No communication with 515 pump on prep system - WKB250347

Article number: 250347

SYMPTOMS

  • No communication with 515 pump on prep system
  • No communication with Waters Pump Control
  • GPIB board present in instruments

ENVIRONMENT

  • 515
  • Autopurification
  • FractionLynx
  • SFO

CAUSE

  • Faulty IEEE cable between PC and SFO on prep system
  • IEEE cable or GPIB board fitted incorrectly

FIX or WORKAROUND

  1. Reseat GPIB board in the PC and IEEE cable between PC and SFO.
  2. Replace the IEEE cable (WAT200491).
  3. Replace the GPIB board (see wkb19969).