No communication with 515 pump on prep system - WKB250347
Article number: 250347
SYMPTOMS
- No communication with 515 pump on prep system
- No communication with Waters Pump Control
- GPIB board present in instruments
ENVIRONMENT
- 515
- Autopurification
- FractionLynx
- SFO
CAUSE
- Faulty IEEE cable between PC and SFO on prep system
- IEEE cable or GPIB board fitted incorrectly
FIX or WORKAROUND
- Reseat GPIB board in the PC and IEEE cable between PC and SFO.
- Replace the IEEE cable (WAT200491).
- Replace the GPIB board (see wkb19969).