Inconsistent injection volumes on 2767 Sample Manager - WKB245886

Article number: 245886

SYMPTOMS

  • Inconsistent injection volumes on 2767 Sample Manager 

ENVIRONMENT

  • MassLynx
  • 2767 Sample Manager 
  • 3767 Sample Manager
  • FractionLynx

CAUSE

Maximum recovery HPLC vials were used and sample probe was touching bottom of vial, causing blockage.

FIX or WORKAROUND

Adjust the Z depth for the sample vial.

ADDITIONAL INFORMATION

To adjust the needle depth into the sample vial, go to Inlet Method editor > Waters2767 > Bed Layout > Modify Bed Layout.

Select each sample rack in turn and adjust the Z height (each value = 1/10 mm). To increase the depth into the vial, increase the value.

clipboard_ef430ad7c3590118b822ffdf0433bde08.png

clipboard_e2916cffe84eaf193d1e93055a8868191.png

id245886, 2767, 2767SFC, 3767