Inconsistent injection volumes on 2767 Sample Manager - WKB245886
Article number: 245886
SYMPTOMS
- Inconsistent injection volumes on 2767 Sample Manager
ENVIRONMENT
- MassLynx
- 2767 Sample Manager
- 3767 Sample Manager
- FractionLynx
CAUSE
Maximum recovery HPLC vials were used and sample probe was touching bottom of vial, causing blockage.
FIX or WORKAROUND
Adjust the Z depth for the sample vial.
ADDITIONAL INFORMATION
To adjust the needle depth into the sample vial, go to Inlet Method editor > Waters2767 > Bed Layout > Modify Bed Layout.
Select each sample rack in turn and adjust the Z height (each value = 1/10 mm). To increase the depth into the vial, increase the value.
id245886, 2767, 2767SFC, 3767