BGM reports an API nitrogen gas failure error in ChromScope system messages - WKB242965

Article number: 242965

SYMPTOMS

  • BGM reports an API nitrogen gas failure error in ChromScope system messages

ENVIRONMENT

  • ChromScope
  • Prep 150 LC System
  • ACQUITY QDa Mass Detector

CAUSE

  • ACQUITY QDa Mass Detector lost nitrogen pressure
  • ACQUITY QDa Mass Detector is not in Operate but flow is started

FIX or WORKAROUND

  1. Reset ACQUITY QDa Mass Detector to clear the API nitrogen gas failure error.
  2. Reset the BGM pump via the Diagnostics menu:Diag Reset.PNG
  3. Reset system alarms through ChromScope:clipboard_ec462527d0d6a947d8ee2b7cfbf6a2408.png

ADDITIONAL INFORMATION

 