Fraction collection is not triggered correctly - WKB235970

Article number: 235970

SYMPTOMS

  • Fraction collection is not being triggered correctly 

  • Fraction simulator suggests collection is completed successfully

  • Fraction collection seems very random

ENVIRONMENT

  • Autopurification
  • FractionLynx
  • MassLynx

CAUSE

Corrupted project or method in MassLynx

FIX or WORKAROUND

  1. Close all software and reboot the PC.
  2. Create a new project using the default project as template.
  3. Recreate all methods and the fraction file.

ADDITIONAL INFORMATION