2767 Sample Manager sample probe does not go down into vial - WKB22556

Article number: 22556

SYMPTOMS

  • 2767 Sample Manager sample probe goes to the correct vial location but does not go down into vial
  • No error message
  • Sample probe goes to the inject port but does not go down
  • Z position recalibrated

ENVIRONMENT

  • 2767 Sample Manager
  • MassLynx

CAUSE

The Reference Calibrated check box is cleared.

FIX or WORKAROUND

In the inlet editor:

  1. Select Waters2767 > Reference Position.
  2. Select the Reference Calibrated check box.
  3. Click OK.

Note: To adjust the x, y, or z coordinates, click image.png.

ADDITIONAL INFORMATION