Low response observed at lower wavelengths with 2998 Photodiode Array Detector - WKB203692

Article number: 203692

SYMPTOMS

  • Low response observed at lower wavelengths with 2998 Photodiode Array Detector

ENVIRONMENT

  • MassLynx
  • 2998
  • AutoPurification

CAUSE

A 1% formic acid solution was prepared as the MS make-up solvent, which saturated the 2998 detector signal at lower wavelengths.

FIX or WORKAROUND

Prepare a 0.01% - 0.1% formic acid solution as the MS make-up solvent.

ADDITIONAL INFORMATION

 