Bad chromatography peaks with ACQUITY UPC2 system - WKB193176
Article number: 193176
SYMPTOMS
- Poor chromatography peak resolution.
- Irreproducibility retention times(PDA)
- The shapes of the chromatography peaks are bad.
ENVIRONMENT
- ACQUITY UPC2
CAUSE
There are various causes.
The following causes should be investigated.
- Pulsating ccBSM pressure
- Leakage at ccBSM
- Clogged ccBSM mixing filter (mixer)
- Degraded ccBSM pump seal and check valve
- Degraded ccBSM CO2 pump head cooler
- Pulsating ABPR pressure
- Degraded ABPR static valve cartridge
- Leakage or blockage somewhere on the system
- Degraded column
- Damaged in-line filter
- Degraded mobile phase solution
- Wrong mobile phase solution
- Degraded sample
- Wrong sample
- Low remaining level in CO2 cylinder
- Main plug of the CO2 cylinder not fully opened
FIX or WORKAROUND
Works that can be done by customers
- Prime ccBSM with methanol for 30 minutes
- Make sure there are no leakage that can be seen or heard
- Perform a leak test of ccBSM (If it fails for 3 times, it should be addressed by a trained Waters engineer)
- Prepare fresh mobile phase
- Prepare fresh sample
- Change the column with a new one
- Check the condition of the CO2 cylinder
If the cause cannot be determined with above, ask a trained Waters engineer to address the problem.
- Replace consumable parts of the ccBSM
- ABPR Static Valve Cartridge
ADDITIONAL INFORMATION
