Z-axis position of the needle changes after 2767 injection - WKB18589

Article number: 18589

SYMPTOMS

  • Z-axis position of the needle changes when homing after injection
  • Position of the needle tip increases or decreases

ENVIRONMENT

  • MassLynx v4.1

CAUSE

The belt of the needle Z axis is loose or damaged.

FIX or WORKAROUND

  • Loose belt - Re-tighten the screws to establish belt tension.
  • Worn belt - Replace the belt.

ADDITIONAL INFORMATION