UPC2 system high-pressure shutdown when performing gradient with high % co-solvent - WKB15645
Article number: 15645
SYMPTOMS
- High pressure and shutdown when using second 30cm Single Zone Column Manager (CM-30S)
- Initial conditions of gradient are stable at 95% CO2 (A) and 5% acetonitrile co-solvent (B)
- At injection, pressure increases as the gradient approaches 40%A and 60%B
- Overpressure occurs during runtime on the second injection
- All eight columns positions in the secondary CM-30S exhibit this overpressure and system shutdown
- Injections made across all columns in the primary CM-30S do not overpressure
- Isocratic method results in good performance across all seven column positions in the primary CM-30S
- Flushing the system overnight, high-pressure shutdowns during injections; re-traced flow paths and checked for blockages
- ABPR health check and diagnostic pass
- CM-30S inlet and outlet tubing new
- Inject, auxiliary, and column switching valve cartridges new
- 2998 flow cell new
- System with fresh solvents used for UPC2 startup testing using a union (no column) and methanol and CO2—pressure and system were fine
- System with the chemists solvent (ACN + 0.2% isopropylamine) and up to 65% co-solvent—the system went over pressure and shut down
- Major spikes observed toward the higher composition of co-solvent
ENVIRONMENT
- ACQUITY UPC2 system:
- ACQUITY UPC2 Binary Solvent Manager (ccBSM)
- ACQUITY UPC2 Sample Manager - Fixed Loop (SM-FL)
- ACQUITY UPC2 30cm Single Zone Column Manager (CM-30S)
- ACQUITY UPC2 Convergence Manager
- ACQUITY 2998 PDA
CAUSE
Co-solvent, mobile phase contamination from topping off solvents
- Co-solvent glass bottle dated November 2015
- Suspect water content in mobile phase contributed to high pressure; composition of co-solvent increased above 50%
FIX or WORKAROUND
Perform a 100% acetonitrile co-solvent flush through the entire UPC2 system:
- Connect each module one at a time in order to wash off any contamination or poor quality solvent from the system components
- Flush each module (approximately one hour) and monitor pressure because high pressure could still occur
ADDITIONAL INFORMATION
id15645, 2998, 2998B, 2998SFC, A-10PDA, eluent, UPC2BSM, UPC2CMA, UPC2CNVMGR, UPC2SMFL