Skip to main content
Waters

UPC2 system high-pressure shutdown when performing gradient with high % co-solvent - WKB15645

Article number: 15645

SYMPTOMS

  • High pressure and shutdown when using second 30cm Single Zone Column Manager (CM-30S)
  • Initial conditions of gradient are stable at 95% CO2 (A) and 5% acetonitrile co-solvent (B)
  • At injection, pressure increases as the gradient approaches 40%A and 60%B
    • Overpressure occurs during runtime on the second injection
    • All eight columns positions in the secondary CM-30S exhibit this overpressure and system shutdown
  • Injections made across all columns in the primary CM-30S do not overpressure
    • Isocratic method results in good performance across all seven column positions in the primary CM-30S
  • Flushing the system overnight, high-pressure shutdowns during injections; re-traced flow paths and checked for blockages  
    • ABPR health check and diagnostic pass  
    • CM-30S inlet and outlet tubing new
    • Inject, auxiliary, and column switching valve cartridges new
    • 2998 flow cell new
  • System with fresh solvents used for UPC2 startup testing using a union (no column) and methanol and CO2—pressure and system were fine
  • System with the chemists solvent (ACN + 0.2% isopropylamine) and up to 65% co-solvent—the system went over pressure and shut down
    • Major spikes observed toward the higher composition of co-solvent

Pressure spikes observed higher composition co-solvent.png

 

ENVIRONMENT

  • ACQUITY UPC2 system:
    • ACQUITY UPC2 Binary Solvent Manager (ccBSM)
    • ACQUITY UPC2 Sample Manager - Fixed Loop (SM-FL)
    • ACQUITY UPC2 30cm Single Zone Column Manager (CM-30S)
    • ACQUITY UPC2 Convergence Manager
    • ACQUITY 2998 PDA

CAUSE

Co-solvent, mobile phase contamination from topping off solvents

  • Co-solvent glass bottle dated November 2015
  • Suspect water content in mobile phase contributed to high pressure; composition of co-solvent increased above 50%

FIX or WORKAROUND

Perform a 100% acetonitrile co-solvent flush through the entire UPC2 system:

  • Connect each module one at a time in order to wash off any contamination or poor quality solvent from the system components
  • Flush each module (approximately one hour) and monitor pressure because high pressure could still occur

ADDITIONAL INFORMATION

id15645, eluent, UPC2BSM, UPC2CMA, UPC2CNVMGR, UPC2SMFL

Not able to find a solution? Click here to request help.