What are the part numbers of the two sizes of 1/8-inch stainless steel tubing for a Bio-Botanical Extraction System? - WKB13512
Article number: 13512
ENVIRONMENT
- Bio-Botanical Extraction System (BBES)
ANSWER
High-pressure tubing for the lines after the pump: 700011275 (Tubing, 1/8"OD X .035"WL. Straight 316SS)
Low-pressure tubing for the lines between the CO2 source (or recycler) and the inlet of the pump: 700011287 (Tubing, 1/8"ODx.020"Wl 5' Straight 316SS)
ADDITIONAL INFORMATION
The difference is the wall thickness. The thinner walled tubing should never be used on the high-pressure side of the CO2 pump because it is not rated for the pressures generated by the P200 pump in this system.
The tubing comes in lengths of 5 feet and must be cut to size.
Tubing cannot be ordered through the Waters website. Please contact Waters Customer Service at 800 252 4752.
id13512, SFEBBC2