Needle hits vials on 2767 - WKB124873

Article number: 124873

SYMPTOMS

  • Needle hits vials

ENVIRONMENT

  • 2767

CAUSE

2767 XYZ is out of alignment.

FIX or WORKAROUND

  1. Check and adjust the reference position for 2767.
  2. Check and adjust the reference position for all plates.

ADDITIONAL INFORMATION

 

id124873, 2767, 2767SFC