2767 Sample Manager injection valve malfunctions - WKB116671
Article number: 116671
SYMPTOMS
- Left Injection Valve: Position Error
- Right Injection Valve: Position Error
ENVIRONMENT
- 2767 Sample Manager
CAUSE
Normal wear and tear, or the valve sensor of the injection valve may be defective.
The valve sensor is vulnerable to impact, and if a strong impact is applied to the injection valve body during the replacement work of the injector valve seal, the valve sensor may have problems and the above symptoms may occur. However, it is not possible to replace only the sensor.
FIX or WORKAROUND
Replace the valve.
ADDITIONAL INFORMATION
Part numbers
- Left injection valve: 700001829 - VALVE, ANALYTICAL, .015 7 PORT, 3POS
- Right injection valve: 700001830 V- ALVE, PREP, .030 7 PORT, 3POS
The valve motors are identical; only the replaceable front parts are different.