Skip to main content
Waters

The temperature of the ThermoCube Chiller goes up during runs on MV-10 or Investigator - WKB102077

Article number: 102077

SYMPTOMS

  • Chiller is set to 3 °C but goes up to about 6°C during run

ENVIRONMENT

  • MV-10
  • Investigator

CAUSE

The CO2 is warming the coolant that is circulating through the pump heads.

FIX or WORKAROUND

This is normal and should not be a concern, unless the temperature of the ThermoCube does not drop back down after the run is over. If there is an issue, it would usually manifest in back pressure not stabilizing and therefore not being able to start injections. 

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.