Y-axis belt slips on 2767 Sample Manager - WKB100315

Article number: 100315

SYMPTOMS

  • Y-axis belt slips on 2767 Sample Manager
  • Grinding noise when moving

ENVIRONMENT

  • MassLynx
  • 2767 Sample Manager

CAUSE

The y axis rail is dirty.

FIX or WORKAROUND

  1. Exit the inlet editor.
  2. Power-off the 2767.
  3. Wipe the Y rails (probe and collector rail) with a lint-free cloth dampened with isopropyl alcohol.
  4. Remove the racks from the bed.
  5. Move the arm to the front-left and down near the bed.
  6. Power-on the 2767.
  7. Open the inlet editor to reset.
    Note: If the arm does not move, select Waters 2767 > Reset Communications in the inlet editor.

ADDITIONAL INFORMATION