FID detector does not ignite - WKB98434
Article number: 98434
SYMPTOMS
- FID not igniting
ENVIRONMENT
- Agilent 6890
- Agilent 7890
CAUSE
- Dirty/oxidized glow plug
- Connection broken between glow plug and GC
FIX or WORKAROUND
- Replace the glow plug (19231-60680).
- Replace the igniter cable assembly (G1531-60680).
ADDITIONAL INFORMATION