Chronos error: Could not parse status.ini file - WKB30579

Article number: 30579

SYMPTOMS

  • Chronos method run halted with error: "Could not parse status.ini file"

ENVIRONMENT

  • HDX-2
  • Chronos 4.5.3
  • MassLynx

CAUSE

MassLynx is not properly configured to write the status.ini files, so the file in the Chronos schedule cannot be read.

FIX or WORKAROUND

  1. In MassLynx, click Options.
  2. On the Options page, change the MassLynx Status refresh rate from 2 to 5 seconds.

Refresh rate.PNG

ADDITIONAL INFORMATION