"Z-axis home h/w fault" with ACQUITY SM-FL - WKB282235

Article number: 282235

SYMPTOMS

  • "Z-axis home h/w fault" error message
  • Error only in certain positions

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Fixed Loop (SM-FL)

CAUSE

The top part of the needle has come away from the fixture/grove and flexed upward, interfering with homing of the sensor flag.

FIX or WORKAROUND

Loop the needle around the front, beneath the plastic hook, instead of behind in the groove. 

ADDITIONAL INFORMATION