Waters Fraction Collector III performs collections randomly. Is there any solution? - WKB24908
Article number: 24908
SYMPTOMS
- Waters Fraction Collector III performs collections randomly.
ENVIRONMENT
- Waters Fraction Collector III
- Simple/Time mode
CAUSE
Poor connection or disconnection of lead wire for 3 WAY valve connection
FIX or WORKAROUND
- On the CONDITIONS screen, check whether 3-WAY VALVE is recognized.
- If it is displayed as "non Attach" which means the valve is not recognized, check the condition of the lead wire and its connection.
- If the lead wire is broken, replace the valve. If the connection is poor, reconnect the wire.