How to cut a piece of stainless tubing and set ferrules - WKB242790

Article number: 242790

OBJECTIVE or GOAL

Cut a stainless steel piece of tubing for HPLC Systems

ENVIRONMENT

  • 510 HPLC pump
  • 515 HPLC Pump
  • 600 MultiSolvent Delivery Pump
  • 717plus Autosampler
  • 1500 Series Column Heater
  • 1515 Isocratic HPLC Pump
  • 1525 Binary HPLC Pump
  • 1525 Micro Binary HPLC Pump
  • 2707 Sample Manager
  • 2777 Sample Manager
  • Alliance HPLC System
  • 2695 Separations Module
  • 2695D Dissolution Separations Module
  • 2795 Separations Module
  • Alliance Series Column Heater
  • Alliance Series Column Heater/Cooler
  • Arc HPLC System
  • 2410 Differential Refractometer
  • 432 Conductivity Detector
  • 2432 Conductivity Detector
  • 2414 Refractive Index Detector
  • 2420 Evaporative Light Scattering Detector
  • 2424 Evaporative Light Scattering Detector
  • 2465 Electrochemical Detector
  • 3465 Electrochemical Detector
  • 2475 Multi-Wavelength Fluorescence Detector
  • 2487 Dual-Wavelength Absorbance Detector
  • 2488 Multichannel Absorbance Detector
  • 2489 UV/Vis Detector
  • 2996 PDA Detector
  • 2998 Photodiode Array Detector

PROCEDURE

  1. Use a 1/16" OD stainless steel tubing cutter (Part Number WAT022384)
  2. Determine the proper 1/16" OD Internal Diameter tubing for use or replacement.
  3. Have the proper ferrules and compression screws available for use.
  4. Have two 5/16-inch wrenches and/or two 1/4-inch wrenches.
  5. Place tubing cutter on the tubing and turn it multiple times till it has been deeply scored.
  6. Use two pliers and flex the scored area until it breaks. Make sure the tubing end is not cut uneven. (Optional Tubing Straightener Part number 700008702)
  7. Place the new ferrules and compression screws on to cut tubing.
  8. Set the ferrules and compression screws by using a union (Part Number WAT097332).
  9. Unscrew the assembled fitting and verify that the length of tubing extending beyond the ferrule is 0.12 inch (3 mm) clipboard_e3d20247ff2cdef43ff2d36862e20be23.png

Not able to find a solution? Click here to request help.