Diverter valve leaks on WFCIII - WKB122939

Article number: 122939

SYMPTOMS

  • Diverter valve leaks even after replacement
  • Solvent recovery is only 95% when collecting at 10–15 mL/min flow

ENVIRONMENT

  • Waters Fraction Collector III

CAUSE

Tubing ID is too small.

FIX or WORKAROUND

Replace with the 0.04-inch ID tubing included in the WFCIII accessory kit (205000131).

ADDITIONAL INFORMATION