Baseline fluctuation on the 2414 in a GPC system caused by tetrahydrofuran stored for an extended period - WKB121309
Article number: 121309
SYMPTOMS
- The baseline fluctuation of the 2414 in a GPC system suddenly increased, while other parameters remained normal
- The pressure of the 1515 pump was slightly higher than previously
- Newly changed tetrahydrofuran solvent
ENVIRONMENT
- GPC/SEC Option
- 1515 Isocratic HPLC Pump
- 2414 Refractive Index Detector
CAUSE
The newly purchased tetrahydrofuran contained a high level of peroxide and water due to an extended storage period, resulting in large baseline fluctuations.
FIX or WORKAROUND
The customer changed suppliers and purchased tetrahydrofuran that was freshly manufactured. After flushing the system and equilibrating the chromatographic system for a long time, the system pressure returned to normal, and the baseline fluctuation after injection returned to the normal range.