Low sensitivity post PM on SYNAPT G2-S - WKB99046

Article number: 99046

SYMPTOMS

Lost 4x sensitivity from immediately after the PM.

ENVIRONMENT

  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS
  • Synapt G2
  • Synapt G2-S
  • Synapt G2-Si

CAUSE

Tuning.

FIX or WORKAROUND

Increase the collector (for example, on a SYNAPT G2-S, raise from 50 to 75).

ADDITIONAL INFORMATION