How to replace a standard probe/capillary assembly on an ACQUITY QDa Mass Detector - WKB95083
Article number: 95083
OBJECTIVE or GOAL
How to replace a standard probe/capillary assembly on an ACQUITY QDa detector.
ENVIRONMENT
- ACQUITY QDa Mass Detector
PROCEDURE
- Stop system flow and put the QDa into Standby mode.
- Unscrew the capillary from the column manager, detector, or ISM outlet.
- Unscrew the connection at the top of the source enclosure cover and remove the capillary from the source enclosure probe guide.
- Carefully drop the replacement capillary through the source enclosure probe guide and tighten the connection to the QDa. Do not force the capillary down.
- Reconnect the capillary to the column Manager, detector, or ISM outlet.
ADDITIONAL INFORMATION
Click here to review the procedure in the ACQUITY QDa Operator guide.
id95083, QDA