MassLynx error: Error starting acquisition. Wrong Ionization mode requested: API mode, Current: ES Mode - WKB9418

Article number: 9418

SYMPTOMS

  • Error starting acquisition. Wrong Ionization mode requested: API mode, Current: ES Mode

ENVIRONMENT

  • MassLynx
  • Xevo G2 Tof

CAUSE

Tune page is in ESI mode but the MS method has both ESI and APCI functions. This requires the instrument to be in ESCi mode.

FIX or WORKAROUND

Place the instrument in ESCi Mode. Select Source - ESCi

ADDITIONAL INFORMATION

 

 