Xevo G2-S QTof fails calibration - WKB92361

Article number: 92361

SYMPTOMS

  • Xevo G2-S QTof fails calibration
  • All peaks flagged with !
  • Happens with both sodium formate and sodium iodide

ENVIRONMENT

  • Xevo G2-S QTof
  • Xevo G2-XS

CAUSE

Veff needs to be adjusted.

FIX or WORKAROUND

  1. Adjust Veff in both positive and negative modes.
  2. Run calibration.

ADDITIONAL INFORMATION