"Detector missing instrument configuration settings" error message in the ACQUITY MS Console for a Xevo TQ-S micro - WKB91820
Article number: 91820
SYMPTOMS
- "Detector missing instrument configuration settings" error message
- Resolution and Mass Scale setting on Engineers tab are all 512 or 2048
- Resolve button on IntelliStart page in ACQUITY MS Console is dimmed (unavailable)
ENVIRONMENT
- Xevo TQ-S micro
CAUSE
Mass Scale and Resolution settings have reset to default values.
FIX or WORKAROUND
- From the MassLynx main page, go to File > Open Data File.
- Select a .raw file that was acquired when the system was last working correctly (do not select OPEN).
- Click on the Experiment button at the bottom for the Data Browser windows to open the Experimental Record.
- Scroll down the file and locate the Engineer Settings section.
- Take note of the values for each of the settings.
- Open the tune page.
- Go to the Engineer Settings.
- Manually specify the values from the Experimental Record into the Engineers settings (must hit Enter for value to be excepted).
- When all values are specified and the presence of a signal is confirmed, close the Engineer settings.
ADDITIONAL INFORMATION
id91820, MLYNX, MLYNXV41, SUPMM, XEVOTQSMIC