How to calibrate to 8000 m/z on a 4K quadrupole - WKB87887
Article number: 87887
OBJECTIVE or GOAL
Calibrate up to 8000 m/z on a QToF system with a 4K RF generator using NaI.
ENVIRONMENT
- Vion IMS QTof
- SYNAPT G2-S
- SYNAPT G2-Si
- SYNAPT XS
- Xevo G2-S
- Xevo G2-XS
PROCEDURE
- Use a manual quad profile with the following settings as a starting point, and adjust accordingly:
Cone Voltage: 150
Quad Profile: 100, 2000, 3940
Dwell Time: 10, 25
Ramp Time: 25, 40
ADDITIONAL INFORMATION
- On a Vion, changing the MS profile does not invalidate the mass and CCS calibrations that are already acquired for up to m/z 1000 and up to m/z 2000.
- After the calibration is successfully performed, return the quad profile to the proper settings for use in data acquisition.
id87887, MALDIG2SI, SYNG2SIHD, VION, XEVOG2XSTF