Skip to main content
Waters

How to calibrate to 8000 m/z on a 4K quadrupole - WKB87887

Article number: 87887

OBJECTIVE or GOAL

Calibrate up to 8000 m/z on a QToF system with a 4K RF generator using NaI.

ENVIRONMENT

  • Vion IMS QTof
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • SYNAPT XS
  • Xevo G2-S
  • Xevo G2-XS

PROCEDURE

  1. Use a manual quad profile with the following settings as a starting point, and adjust accordingly:
    Cone Voltage: 150
    Quad Profile: 100, 2000, 3940
    Dwell Time: 10, 25
    Ramp Time: 25, 40 

ADDITIONAL INFORMATION

- On a Vion, changing the MS profile does not invalidate the mass and CCS calibrations that are already acquired for up to m/z 1000 and up to m/z 2000.

- After the calibration is successfully performed, return the quad profile to the proper settings for use in data acquisition.

Not able to find a solution? Click here to request help.