Skip to main content
Waters

How to manually calibrate the ToF using column bleed ions on APGC - WKB8446

Article number: 8446

OBJECTIVE or GOAL

Use naturally eluting background peaks to provide a calibration for APGC. (The peaks used are column bleed (siloxane) peaks with a plasticizer peak added as a low mass calibrant.)

ENVIRONMENT

  • APGC
  • MassLynx
  • SYNAPT G2
  • Xevo G2-XS
  • Xevo G2-S

PROCEDURE

  1. Open up the MS Console.
     Siloxane 1.png
  2. Browse to IntelliStart, and then select Configure / Reference File Editor.
     Siloxane 2.PNG
  3. Select File > Import.
     Siloxane 3.PNG
  4. Select the Siloxane.ref file. See Additional Information if this file is not present.
     Siloxane 4.PNG
  5. Fill out the new reference compound information as shown below, ensuring that the polarity selected is positive.
     Siloxane 5.PNG
  6. Run a tune page acquisition to get siloxane data under current conditions. Note that the data format must be continuum and the mass range must be the same as the calibration profile that will be created later (50 to 1200 Da is a good general range). The instrument mode (i.e., Sensitivity, Resolution, or High Resolution) musst also match. If calibration points at higher masses (>~600 Da) are required, turning the GC oven temperature up can help get higher-mass siloxanes into the source.
     Siloxane 6.PNG
  7. Open a chromatogram of the data acquired and, with a right-click drag, combine the entire set of data to a single spectrum.
     Siloxane 7.PNG
  8. In the spectrum window, select Process > Automatic Peak Detection. Use the automatic parameter selection when prompted, and press OK.
     Siloxane 8x1.png

    Siloxane 8x2.png
  9. Once the peak detection process has run, click the save button on the right-hand side of the toolbar, and click OK when prompted.
     Siloxane 9x1.PNG

    Siloxane 9x2.PNG
  10. Return to the MS Console and select Calibration Profile Editor from the Configure menu.
     Siloxane 10.png
  11. Select File > New from the calibration profile editor.
     Siloxane 11.PNG
  12. Fill out the details for the calibration profile as shown below, setting the mass range as before and selecting manual calibration. Under positive polarity, select the reference compound from the drop-down menu for the mode in which the data were acquired (“Resolution” in the example below).
     Siloxane 12.png
  13. In the adjacent Raw Data box, click the browse button and select the data from step 6.
     Siloxane 13.png
  14. The calibration profile should appear as below. Select OK when done.
     Siloxane 14.PNG
  15. Returning to the MS Console page, select the Create Calibration check box and click Start.
     Siloxane 15.PNG
  16. Proceed through the following screens as shown, selecting the calibration profile just created.
     Siloxane 16x1.PNG
    Siloxane 16x2.PNG
  17. Manually calibrate the spectrum and click the green check mark in the top-left corner when satisfied.
     Siloxane 17.PNG
  18. A calibration report should appear and the MS Console should report that the calibration was successful. Also, the active calibration should now appear in the bottom right-hand side of the tune page.
    Siloxane 18x1.PNG
    Siloxane 18x2.png

ADDITIONAL INFORMATION

The siloxane calibrant is attached below.

Siloxanes.ref

id8446, APGC, MLYNX, MLYNXV41, SUPMM, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, XEVOG2SQTF, XEVOG2STOF, XEVOG2XSTF

Not able to find a solution? Click here to request help.