QDa Source pressure test leak trip - WKB78106
Article number: 78106
SYMPTOMS
- Source pressure test failed
ENVIRONMENT
- ACQUITY QDa Mass Detector
CAUSE
Nitrogen generator isn’t producing the given volume of gas
FIX or WORKAROUND
- Perform the source pressure test again and set to 20 L/min with the flow rate adjustment nozzle on the nitrogen gas generator as nitrogen gas is flowing, it should be improved.