Article number: 78106

SYMPTOMS

  • Source pressure test failed

ENVIRONMENT

  • ACQUITY QDa Mass Detector

CAUSE

Nitrogen generator isn’t producing the given volume of gas

FIX or WORKAROUND

  1. Perform the source pressure test again and set to 20 L/min with the flow rate adjustment nozzle on the nitrogen gas generator as nitrogen gas is flowing, it should be improved.

ADDITIONAL INFORMATION

 