Communication lost between 2998 PDA detector and MassLynx - WKB72897
Article number: 72897
SYMPTOMS
- Communication lost to 2998 Photodiode Array (PDA) Detector
- Resetting the PDA through the Console results in a spinning donut cursor
ENVIRONMENT
- 2695 Separations Module
- 2998 Photodiode Array Detector
- ACQUITY QDa Mass Detector
- MassLynx
CAUSE
Unknown
FIX or WORKAROUND
- Close MassLynx software.
- Restart the MassLynx PC.
- While PC is restarting, power-off all components, including the QDa.
- Log in to Windows but do not start MassLynx.
- Power-on all components. Wait for all components to complete initialization.
- Open MassLynx.