Gas leak around probe and probe adjuster assembly on Xevo MS - WKB7204
Article number: 7204
SYMPTOMS
- Gas leaks from probe adjuster assembly
- Collar that holds probe to adjuster assembly is loose
- Source pressure test failed error message
- Source pressure test leak warning
ENVIRONMENT
- Xevo TQD
- SQD2
- Xevo TQ-S micro
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- Vion IMS Qtof
CAUSE
Collar of probe is not secured to the probe adjuster assembly.
FIX or WORKAROUND
- Rotating it fully clockwise, tighten the probe collar to secure it to the adjuster assembly
- Open and close the source door to rerun the source pressure test
ADDITIONAL INFORMATION