Gas leak around probe and probe adjuster assembly, and source pressure test error on Xevo TQD - WKB7204

Article number: 7204

SYMPTOMS

  • Gas leaks from probe adjuster assembly
  • Collar that holds probe to adjuster assembly is loose
  • Source pressure test failed error message

ENVIRONMENT

  • Xevo TQD
  • SQD2

CAUSE

Collar of probe is not secured to the probe adjuster assembly.

FIX or WORKAROUND

  1. Rotating it fully clockwise, tighten the probe collar to secure it to the adjuster assembly.
  2. Clear the error in the ACQUITY MS console.
  3. Open and close the source door.

ADDITIONAL INFORMATION

 