Xevo G2-XS "Source leak" reported in UNIFI - WKB69370

Article number: 69370

SYMPTOMS

  • Red light for the status indication
  • The instrument error log shows "Source leak", but there is NO source leak present
  • When the source pressure test is run, it passes the first section but never completely finishes (there is no loud hiss at the end)

ENVIRONMENT

  • UNIFI
  • Xevo G2-XS

CAUSE

A problem with the bottle waste trap assembly 

FIX or WORKAROUND

Replace the bottle waste trap assembly (700004606).

ADDITIONAL INFORMATION

 

id69370, XEVOG2XSTF