Low signal in ESI negative ion mode - WKB64035
Article number: 64035
SYMPTOMS
- Signal lower than expected for analytical methods in ESI negative ion mode
- QDa - internal calibration signal looks normal
ENVIRONMENT
- ACQUITY QDa
CAUSE
Higher than necessary formic acid concentration in mobile phase
FIX or WORKAROUND
Reduce the concentration of formic acid in the mobile phase to the lowest level workable for the method. A concentration of 0.01% formic acid (v/v) is considered appropriate.
ADDITIONAL INFORMATION
Formic acid suppresses ionization in ESI negative ion mode. In positive ion mode, it can enhance signal.
id64035, Electrospray, eluent, QDA