Xevo vents after cleaning the sample cone and cone gas nozzle - WKB6020
Article number: 6020
SYMPTOMS
- System vents after cleaning the sample cone and cone gas nozzle
- Pumps down, but turbo speed reaches only 50-60 %, and then it vents
- With isolation valve closed, pumps down properly
ENVIRONMENT
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQD
- MassLynx
- Xevo G2-XS QTof
CAUSE
Damaged sample cone O-ring
FIX or WORKAROUND
Replace the sample cone O-ring (700004321 - O-ring fluoroelastomer size 10 I/D x 1.5 C/S).
ADDITIONAL INFORMATION