QDa turbo speed does not exceed 5% - WKB58573

Article number: 58573

SYMPTOMS

  • Turbo does not spin more than 5% when unit is powered-on
  • The rotary pump does not start up

ENVIRONMENT

  • ACQUITY QDa Mass Detector

CAUSE

The rotary pump power switch is off.

FIX or WORKAROUND

  1. Turn off the QDa.
  2. Switch the rotary pump power switch to ON.
  3. Power-up the QDa.