Mass spectometer starts the acquisition before the inlet system can start the injection cycle - WKB57938
Article number: 57938
SYMPTOMS
- The mass spectrometer starts acquiring immediately upon starting the sample list
- Inlet method does not run
- No injection cycle
- Found: Event in 1 is selected in the Inlet Configuration, Events and Triggering
ENVIRONMENT
- ZQ 2000
- ACQUITY UPLC
- ACQUITY UPLC Sample Manager
- MassLynx 4.1
CAUSE
Poorly connected contact closure cable on the back of the ACQUITY ManagerUPLC Sample
FIX or WORKAROUND
Properly connect the contact closure cable in the terminal block on the back of the ACQUITY Sample Manager.
ADDITIONAL INFORMATION
id57938, UPBINARY, ZQ2000, ZQ2000SFC, ZQ4000