Mass spectrometer signal is low, capillary voltage is not active - WKB52512
Article number: 52512
SYMPTOMS
- Nonexistent or low signal from MS
- Capillary readback shows voltage does not reflect set point
ENVIRONMENT
- Xevo TQD
- SQ Detector 2
CAUSE
Power outages, loss of communications, and crash venting can all cause this issue.
FIX or WORKAROUND
- Adjust the capillary voltage slider, or enter a different set point for the capillary voltage.
- Once the readback responds, set the capillary voltage back to the previous value.
ADDITIONAL INFORMATION