Poor mass accuracy on Xevo G2-S - WKB4904
Article number: 4904
SYMPTOMS
- Calibration setup completes successfully, but mass accuracy is poor
- No information about calibration in Experimental Record
- Poor mass accuracy on ToF data
- Poor mass accuracy on IMS data, but great mass accuracy on ToF data on Synapt MS
- Experiment file shows that no calibration information is properly applied to acquired data.
- Calibration file is observed to be loaded as per the MS tune page and the Intellistart console
ENVIRONMENT
- MassLynx 4.1, 4.2
- Windows 7
- Windows 10
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- SYNAPT XS
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
CAUSE
Calibration profile name contains spaces or special characters.
FIX or WORKAROUND
Reinstall MassLynx.
ADDITIONAL INFORMATION
Calibration not applied properly Calibration applied properly