Constant or Intermittent communication and dropouts caused by faulty EPC - WKB4863

Article number: 4863

SYMPTOMS

  • Intermittent communication with MS or no communicaions at all
  • Random dropouts or system unresponsiveness seen while idle, in Standby, or during acquisitions

Errors seen may include any the following at different times:

  • Cannot communicate to EPC via Hyperterminal
  • Cannot ping IP address of EPC
  • NetEvaluator shows red / no connection to the MS, but the inlets are okay
  • LED light on EPC is red
  • Error message: "Probe not connected"
  • On board diagnostic page, the RF Barrier PCB is not visible
  • Error "lost connection to embedded PC" and must perform a full reboot of PC and MS

ENVIRONMENT

  • Xevo TQ Absolute
  • Xevo TQ cronos
  • Xevo TQ micro
  • Xevo TQ-XS
  • Xevo TQ-S
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-GC
  • SQ Detector 2
  • MassLynx

CAUSE

Faulty EPC

FIX or WORKAROUND

  1. Replace the EPC 
  2. Check which MassLynx version supports the new EPC. (Select your instrument from this article  WKB554)

ADDITIONAL INFORMATION

 