Constant or Intermittent communication and dropouts caused by faulty EPC - WKB4863
Article number: 4863
SYMPTOMS
- Intermittent communication with MS or no communicaions at all
- Random dropouts or system unresponsiveness seen while idle, in Standby, or during acquisitions
Errors seen may include any the following at different times:
- Cannot communicate to EPC via Hyperterminal
- Cannot ping IP address of EPC
- NetEvaluator shows red / no connection to the MS, but the inlets are okay
- LED light on EPC is red
- Error message: "Probe not connected"
- On board diagnostic page, the RF Barrier PCB is not visible
- Error "lost connection to embedded PC" and must perform a full reboot of PC and MS
ENVIRONMENT
- Xevo TQ Absolute
- Xevo TQ cronos
- Xevo TQ micro
- Xevo TQ-XS
- Xevo TQ-S
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-GC
- SQ Detector 2
- MassLynx
CAUSE
Faulty EPC
FIX or WORKAROUND
- Replace the EPC
- Check which MassLynx version supports the new EPC. (Select your instrument from this article WKB554)