Low sensitivity for opiates assay on Xevo TQD - WKB46364

Article number: 46364

SYMPTOMS

  • Opiates assay shows decreased response for all transitions
  • Cleaning the source block assembly and source cone does not improve the situation
  • Setup solution infusion tests shows normal response for MS1 and MS2

ENVIRONMENT

  • Xevo TQD

CAUSE

Dirty or compromised ESI probe capillary

FIX or WORKAROUND

Rebuild the ESI probe and replace the capillary.

ADDITIONAL INFORMATION

 

id46364, Electrospray, XEVOTQD, XEVOTQDIVD