"Turbo controller fault" error on mass spectrometer - WKB4358

Article number: 4358

SYMPTOMS

  • Turbo controller fault error message
  • Instrument did not vent

ENVIRONMENT

  • SQ Detector 2
  • TQD
  • Xevo TQD
  • SQD

CAUSE

The roughing pump is out of oil.

FIX or WORKAROUND

Refill the oil in the roughing pump.

ADDITIONAL INFORMATION