How do I enable a Startup/Shutdown method in UNIFI? - WKB3441
Article number: 3441
OBJECTIVE or GOAL
To enable a Startup/Shutdown method in UNIFI.
ENVIRONMENT
- Vion
- Xevo G2-XS QTof
- Xevo G2-S QTof
- Xevo G2 QTof
- Xevo TQ
- Xevo TQD
- Xevo TQ-S
- I-Class
- H-Class
- M-Class
- All versions of UNIFI up to and including:
- UNIFI V1.8.0
- UNIFI V1.8.1
- UNIFI V1.8.2
PROCEDURE
- Go to Administration > Device Management > Instrument System and Modify instrument system.
- The Startup/Shutdown tab is now available to select.
- In Settings, choose the method you want to apply.
- Select Enable Startup/Shutdown and the relevant time.
- You can schedule the event, if necessary.
ADDITIONAL INFORMATION
This procedure assumes that a startup/shutdown method has already been created.
For information on how to modify a startup or shutdown method in UNIFI, select this link.