"MS Error" in the ACQUITY MS Console for a Xevo TQ-MS - WKB3114
Article number: 3114
SYMPTOMS
- System Status is green
- Error message "MS Error"
- Cannot turn on API gas but can turn on collision gas
ENVIRONMENT
- Xevo TQ-MS
- MassLynx 4.1
CAUSE
Xevo TQ-MS has only partial communications with the MassLynx PC.
FIX or WORKAROUND
- Close MassLynx.
- Restart the MassLynx PC. Log in to Windows but do not start MassLynx.
- Restart the embedded computer (EPC) by pushing the reset button on the front of instrument, inside the small hole.
- Wait approximately three minutes for the EPC to restart.
- Open MassLynx and the MS Console.
ADDITIONAL INFORMATION
id3114, comms, MLYNX, MLYNXV41, SUPMM