MS Quan does not show data after processing - WKB290730
Article number: 290730
SYMPTOMS
- After processing data, MS Quan shows for some components only results between RT 0 and 1 minute
- LCMS Toolkit shows all data
ENVIRONMENT
- waters_connect
- MS Quan
CAUSE
Two different acquisition methods are used with slightly different optimized MRM transitions. The differences were in hundreds. This can be caused by copying an MS Method from one workstation to another.
The MS Quan software uses not the function numbers but the MRM transition to link the correct channel to the component.
FIX or WORKAROUND
- Ensure that the masses for the MRM transitions in the MS Method are the same as in the processing method.
- Use the same method on both systems.