How to remove, disassemble, clean, and refit the source components to the RADIAN ASAP Direct Mass Detector - WKB288927

Article number: 288927

OBJECTIVE or GOAL

How to remove, disassemble, clean, and refit the source components to the Radian ASAP Mass Detector.

INCLUDES:

Replacing / cleaning corona pin, source enclosure pumping block o-ring removal, removal, disassembly and cleaning of the ion guide assembly (including differential aperture disc) and reassembly of the ion guide and replacing the source aperture disc.

ENVIRONMENT

  • RADIAN ASAP Direct Mass Detector

PROCEDURE

  1. Click here to review procedures within the RADIAN ASAP Direct Mass Detector Operators Guide

ADDITIONAL INFORMATION

VIDEO on Waters.com

How to remove, disassemble, clean, and refit the source components to the Radian ASAP mass detector.

Removal and disassembly of source components 00:00:40 - 00:06:52 mins

Cleaning source components 00:06:52 - 00:11:12 mins

Refitting source components 00:11:12 - 00:15:57 mins

Not able to find a solution? Click here to request help.