Communication with EPC or ADC is intermittent - WKB28228
Article number: 28228
SYMPTOMS
- The system communication is intermittently lost and the system requires several restart attempts to reestablish communication.
- Network diagnostics between PC and EPC can be established (Telnet EPC works).
- Sonar upgrade not successful, if applicable.
- May be unable to communicate with the ADC Lite PCB, (see below).
- Copy of error seen on the restart script captured from Hyperterminal application:
[203621047379] (tShell0): Creating ADC data acquisition class
[279624812030] (tShell0): Status: ERROR: Failed to connect to ADC module.
[...]
sync_ADC_files_to_ADC
ptimizationRoutAdcFtpTransfer: Directory /ata0/Adc already exists
ines:Info:OpAdcUpdate: failed to send EV_ADC_SET_FILE_UTILS_DIR
ADC files to ADC ERROR.
value = -1 = 0xffffffff - ADC PCB shows LEDs status OK on front panel during full electronics restart as per technical reference sheet 715003767.
ENVIRONMENT
- SYNAPT G2-Si
- Xevo G2-XS
- MassLynx
CAUSE
Unknown. Probable fault during the loading of MassLynx or the setup of the acquisition host PC.
FIX or WORKAROUND
- Re-image the Workstation.
- Install MassLynx.
ADDITIONAL INFORMATION
MS instrument is controlled successfully with different PC. Able to see all of board with the right firmware versions (for example, in the case of a SONAR upgrade: Lens Control PCB firmware version "W" or later, and ADC firmware version "I" or later).
id28228, comms, MALDIG2SI, MLYNX, MLYNXV41, PuTTY, SUPMM, SYNG2SIHD, XEVOG2XSTF