Article number: 267827

SYMPTOMS

  • Leak in the diverter valve - pump to diverter valve tubing connection
  • Diverter Rotor/Stator Kit (part 700009465) was replaced
  • Replacing the two-piece connector in the pump to diverter valve tubing with a long fitting resolves the leak, but the fluidics pump judders, making a loud noise

Screenshot_2023-08-01-20-50-31-36_99c04817c0de5652397fc8b56c3b3817.jpg

ENVIRONMENT

  • Xevo TQD
  • Xevo TQ-S micro 

CAUSE

The waste line in the diverter valve is partially blocked or kinked, creating extra pressure and causing a leak in the weakest connection point.

FIX or WORKAROUND

Unblock the waste line or replace the waste tubing. 

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.