Very low signal in the mass spec chromatogram for the SQ Detector 2 - WKB262578

Article number: 262578

SYMPTOMS

  • Very low signal for the MS chromatogram; no peak detected
  • Signal for the PDA chromatogram is normal; all expected peaks detected
  • FOUND: Capillary Voltage readback is 0.09 kV for a setting of 3.0 kV
  • FOUND: Phospor and Dynode voltage readbacks are both 0 kV
  • Rebooting EPC does not help

ENVIRONMENT

  • SQ Detector 2
  • MassLynx

CAUSE

Unknown communication or electronics glitch

FIX or WORKAROUND

  1. From the Tune page, vent the SQD2 (Vacuum > Vent).
  2. Close MassLynx and the ACQUITY MS Console.
  3. Power-off the SQD2, and wait about 1 minute.
  4. Power-on the SQD2, and wait for the chime at the end of the boot sequence.
  5. Open MassLynx and the MS Tune page.
  6. From the Tune page, start the vacuum system for the SQD2 (Vacuum > Pump).
  7. When the turbo pumps reach 100%, turn on the API gas and put the SQD2 into Operate.
  8. Verify that the readbacks for the Capillary, Dynode, and Phosphor are now correct.

ADDITIONAL INFORMATION