Low sensitivity with QDa - WKB255886
Article number: 255886
SYMPTOMS
- Low sensitivity with QDa
- Calibrant works as expected
ENVIRONMENT
- ACQUITY QDa Mass Detector
CAUSE
Dirty ion optics
FIX or WORKAROUND
- Clean/replace front end with refurb (EEE010377).
- Clean segmented quad.
ADDITIONAL INFORMATION
Click here to review the source cleaning procedures within the ACQUITY QDa Detector Overview and Maintenance Guide section 3.6.
Follow the appropriate section of the video on www.waters.com
Note that the video follows the RADIAN ASAP procedures, but removing, disassembling, cleaning, and refitting the source components is identical to the QDa Detector.
Removal and disassembly of source components 00:02:34 - 00:06:52 mins
Cleaning source components 00:06:52 - 00:11:12 mins
Refitting source components 00:11:12 - 00:15:23 mins
id255886, QDA