Tof vacuum out of range - WKB25548

Article number: 25548

SYMPTOMS

  • System fails to go into Operate mode because Tof vacuum is out of range, either low (1e-8 mBar) or <1e-4 mBar.

ENVIRONMENT

  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Xevo QTof
  • Xevo Tof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof

CAUSE

Bad Tof Penning gauge body

FIX or WORKAROUND

Replace the Penning gauge body, BODY TUBE SPARES ASSY AIM, part number 6061138.